Description
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
Details
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N
Affected Packages
| Software | From version | Fixed in |
|---|---|---|
| battery-pack-sp-with-wifi-firmware | — | 054u000092 |
References
Free Vulnerability Check
Is your WordPress site affected?
BotEraser helps you identify potentially vulnerable plugins and themes by checking your installation against known CVE records.
Scan My Site Free →No credit card required · Results in minutes
ⓘ Data Notice: The information presented above has been compiled from publicly available internet sources. Boteraser aggregates this data solely for informational purposes and does not independently classify, evaluate, or endorse any findings about the vulnerabilities listed. The accuracy and completeness of this information is the sole responsibility of the original publishers. Boteraser and its operators accept no liability for any decisions made based on this data.